中文版
Home   About   News   Products   Contact
 
Products
 
Reflow Oven (1)
PCB Insertion Loss Measurement (2)
PCB Reliability Test (5)
Bentham Optical Measurement (12)
Compliance West Tester (14)
Resistance to Abnormal Heat and to Fire (0)
Joint finger and mechanism secure testing apparatus (0)
Waterproof and dust-proof testing apparatus (0)
IEC/UL electric secure testing apparatus (0)

Hipot Tester HT-20KVP AC  
 


HT-20KVP AC Hipot Tester

 

For use in production line testing of high voltage systems, components, product design and type testing.

 

The HT-20KVP-AC is a 20,000 Volts AC only hipot tester, can deliver up to 5mA at 16 KV and 1 mA at 20KV, Test

time, Voltage ramp and Leakage limit settings can be modified in the front panel. Voltage Ramp, Arc detection and

Test timer are defeatable, one year warranty, calibration certificate to ANSI Z540 included.

 

 

Specifications

Voltage Test Type AC
Voltage Range

0-20,000 VAC

Current Range

5.0 mA at 16,000V

1.0 mA at 20,000V

Input Voltage Requirements

120V, 60Hz

Fuse Requirements 120/230V, 2A Slow Blow

Arc Detect

Independent High Frequency  circuit

Leakage Detect

Independent Low Frequency  circuit
Test Time Range 1 - 60 Seconds Defeatable
Voltage Meter Accuracy +/- 0.1 KV full scale
Current Meter Accuracy +/- 0.1 mA full scale
Duty Cycle 100%
Discharge Time < 7mS 
Transformer Ouput < 500 VAC
Display 4 digits LED display
Test Indicator LED indicators
Failure Indicator LED indicators and buzzer
Programable No
PLC Interface No
TestLink No
TestMinder No
Test Defeat Test Time, Voltage Ramp and Arc Detection
Operation Temperature 15 - 40 ℃
Relative Humidity Range 0 - 90% Non-condensing
Dimensions W 17 x H 11 x D 17 (inch)
Weight 45 Lbs

 

Features

  • AC output up to 20,000 V
  • Leakage current up to 5.0 mA.
  • Microcontroller technologies, one button test, easy to setup, LEDs indicates test in progress. Upon failures buzzer sounds and testing is terminated.
  • Output voltage, Leakage current and test duration are controlled in the front panel.
  • Test timer is defeatable.
  • HV connector output with a quarter turn twist lock.
  • Designed for product development and laboratory use.
  • One year calibration cycle.
  • Main shutdown switch in front panel for safety.

 

 
       
 

 